Assertion

AssertionId
LAB27-007 Testable
predicate
An analyzer supporting the "Query by Tray" Option SHALL be capable of sending a query identifying the position of the specimen in the tray. The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-6, QPD-7 and QPD-8. The first four are Mandatory, the latter is Required if Available. QPD-6 identifies the tray and QPD-7 is the position in the tray
Prescription level
Mandatory / Required / Shall
Page
171
Section
3.Q.5.4
Status
validated
Last changed
3/7/16 2:23:29 PM by mpattillo
Comment

Applies to

Audit Messages

Integration Profiles

Transactions

Standards

Covered by

Tests

Test Steps

Rules

MBV

Covered by (Deprecated)

Tests

Test Steps

Rules

Document name
Provenance
Revision
Action
IHE PaLM Technical Framework Supplement - Laboratory Analytical Workflow (LAW)IHE1.5