LAB28LAB28-003validatedTestable 0 1 The Analyzer SHALL respond to the AWOS Broadcast message with an ORL^O24^ORL_O42 message compliant with table 3.R.5.2-2.180Section 3.R.5.13/7/16 2:23:29 PM by mpattillo
LAB28LAB28-004validatedTestable 0 1 If the OML message contains the Order Control Code NW, as part of the ORL^O34 response, the Analyzer SHALL transmit either Notification of request accepted or Unable to accept order/service for each AWOS contained in the OML message.185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB28LAB28-005validatedTestable 0 1 If the OML message contains the Order Control Code CA, as part of the ORL^O34 response, the Analyzer SHALL transmit either Cancel as requested or Unable to cancel for each AWOS contained in the OML message.185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB28LAB28-006validatedTestable 0 1 If the Analyzer Manager retransmits an AWOS or reuses an AWOS ID, the Analyzer SHALL reject the AWOS request by responding with Unable to accept order/service. 185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB28LAB28-008validatedTestable 0 5 A Negative Query Response for an unrecognized specimen is considered an unexpected situation, the Analyzer SHALL respond with an Application Acknowledgement: Reject in the ORL^O34 by setting MSA-1 to AR and reporting the appropriate specimen container values in the ERR segment.185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB28LAB28-009validatedTestable 0 0 In response to a negative query response, the Analyzer shall send an ORL^OO34^ORL_O42 message containing only the MSH and MSA segments.191Section 3.R.5.33/7/16 2:23:29 PM by mpattillo