LAB27 | LAB27-001 | validated | Testable |
0
|
5
| | After the Analyzer working in query mode recognizes one or more specimens, the Analyzer sends a WOS Query Message (QBP^Q11^QBP_Q11) for each specimen to the Analyzer Manager. | 168 | Section 3.Q.5 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB27 | LAB27-005 | validated | Testable |
0
|
1
| | An analyzer supporting the "Query All" Option SHALL be capable of sending a query for all pending work by using the query name WOS_ALL in field QPD-1. The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2. | 171 | Section 3.Q.5.4 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB27 | LAB27-006 | validated | Testable |
0
|
1
| | An analyzer supporting the "Query by Rack" Option SHALL be capable of sending a query identifying the position of the specimen in the rack (carrier). The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-4, QPD-5 and QPD-8. The first four are Mandatory, the latter is Required if Available. QPD-4 identifies the rack and QPD-5 is the position in the rack | 171 | Section 3.Q.5.4 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB27 | LAB27-007 | validated | Testable |
0
|
1
| | An analyzer supporting the "Query by Tray" Option SHALL be capable of sending a query identifying the position of the specimen in the tray. The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-6, QPD-7 and QPD-8. The first four are Mandatory, the latter is Required if Available. QPD-6 identifies the tray and QPD-7 is the position in the tray | 171 | Section 3.Q.5.4 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB27 | LAB27-008 | validated | Testable |
0
|
1
| | An analyzer supporting the "Query by Isolate" Option SHALL be capable of sending a query identifying an isolate within the primary (parent) specimen. The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-3, and QPD-9. All four are Mandatory. QPD-3 identifies the isolate and QPD-9 identifies the primary specimen from which the microorganism was isolated. | 171 | Section 3.Q.5.4 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB27 | LAB27-009 | validated | Testable |
0
|
1
| | An analyzer supporting the "Bi-directional Communication" Option SHALL be capable of sending a query identifying the container. The only fields populated in the QPD segment SHALL be QPD-1, QPD-2, and QPD-3 (see Table X.5.1). | 171 | Section 3.Q.5.4 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB28 | LAB28-003 | validated | Testable |
0
|
1
| | The Analyzer SHALL respond to the AWOS Broadcast message with an ORL^O24^ORL_O42 message compliant with table 3.R.5.2-2. | 180 | Section 3.R.5.1 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB28 | LAB28-004 | validated | Testable |
0
|
1
| | If the OML message contains the Order Control Code NW, as part of the ORL^O34 response, the Analyzer SHALL transmit either Notification of request accepted or Unable to accept order/service for each AWOS contained in the OML message. | 185 | Section 3.R.5.3 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB28 | LAB28-005 | validated | Testable |
0
|
1
| | If the OML message contains the Order Control Code CA, as part of the ORL^O34 response, the Analyzer SHALL transmit either Cancel as requested or Unable to cancel for each AWOS contained in the OML message. | 185 | Section 3.R.5.3 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB28 | LAB28-006 | validated | Testable |
0
|
1
| | If the Analyzer Manager retransmits an AWOS or reuses an AWOS ID, the Analyzer SHALL reject the AWOS request by responding with Unable to accept order/service. | 185 | Section 3.R.5.3 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB28 | LAB28-008 | validated | Testable |
0
|
5
| | A Negative Query Response for an unrecognized specimen is considered an unexpected situation, the Analyzer SHALL respond with an Application Acknowledgement: Reject in the ORL^O34 by setting MSA-1 to AR and reporting the appropriate specimen container values in the ERR segment. | 185 | Section 3.R.5.3 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB28 | LAB28-009 | validated | Testable |
0
|
0
| | In response to a negative query response, the Analyzer shall send an ORL^OO34^ORL_O42 message containing only the MSH and MSA segments. | 191 | Section 3.R.5.3 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB29 | LAB29-001 | validated | Testable |
0
|
1
| | The Analyzer SHALL notify the Analyzer Manager of the test results using the OUL^R22^OUL_R22 message. | 191 | Section 3.Y.5.3 | 11/18/16 5:24:28 PM by admin |
|
LAB29 | LAB29-005 | validated | Testable |
0
|
0
| | When more than one OBX segment is used to send related observations, different codes must be used in OBX-3 for each observation so that OBX-3 and OBX-4 are unique across OBXs. | 92 | Section W.2.5.2 | 3/7/16 2:23:29 PM by mpattillo |
|
LAB29 | LAB29-006 | validated | Testable |
0
|
0
| | The value of OBX-11 field across LAB-29 messages shall respect the Observation Result Status lifecycle for an Observation Run. | 135 | Figure W.3.6-1 | 3/7/16 2:23:29 PM by mpattillo |
|
LAW | LAW-001 | validated | Testable |
0
|
1
| | Upon notification of AWOS completion by one of the Analyzers who transmits back the results, the Analyzer Manager shall cancel the other redundant AWOS awaiting execution on the other Analyzers. | 32 | Section X.2.1 | 3/7/16 9:59:00 AM by aberge |
|
LAW | LAW-002 | validated | Testable |
0
|
0
| | In response to the AWOS broadcast, the Analyzer must notify the Analyzer Manager with its intent to accept or reject an AWOS by sending an ORL^O34^ORL_O42 message. | 32 | Section X.2.1.1 | 3/7/16 9:59:37 AM by aberge |
|
LAW | LAW-004 | validated | Testable |
0
|
0
| | If an Analyzer cannot cancel the AWOS, it should notify the Analyzer Manager that it is unable to cancel (using ORL^O34^ORL_O42 response message). | 34 | Section X.21.1 | 3/7/16 2:23:29 PM by mpattillo |
|
LAW | LAW-005 | validated | Testable |
0
|
0
| | The AWOS ID is never entered manually on the Analyzer. It can only be obtained via a message from the Analyzer Manager. Therefore, when the AWOS is entered at the Analyzer, the AWOS shall have a value of NULL ("") for the null AWOS ID. | 36 | Section X.2.3 | 3/7/16 2:39:37 PM by aberge |
|
LAW | LAW-006 | validated | Testable |
0
|
1
| | The Analyzer Manager must support receiving AWOS test results reported over a period of time in addition to receiving multiple test results for an AWOS at the same time. | 36 | Section X.2.3 | 3/7/16 10:13:57 AM by aberge |
|