Standard details

Standard details

Keyword: HL7v2.6

Name: HL7 Version 2.6

Version: 2.6

HL7v2
Assertion Id
Description
DEC-Vol2-001 The Communicate PCD Data [PCD-01] transaction references the following chapters from the HL7 Version 2.6 standard: Chapter 7 Observation Reporting
DEC-Vol2-004 The static definition for the PCD-01 Communicate PCD Data message is defined in the unnamed table in this section of the doument
DEC-Vol2-005 The PCD-01 message structure differs from the basic HL7 version 2.6 by allowing for the appearance of PRT segments, a segment new in HL7 version 2.7, in certain locations.
DEC-Vol2-006 The period for periodic reporting shall be several times a minute for high acuity reports and maximum interval of once every 24 hours, with a typical interval of 1 minute.
DEC-Vol2-007 Upon receipt of a PCD-01 message the DOC actor validates the message and shall respond with an acccept acknowledgement message as defined in Appendix G.11 Acknowledgement Modes.
DEC-Vol2-008 For PCD-01 messages MSH-9 Message Type shall contain ORU^R01^ORU_R01.
DEC-Vol2-009 Device to Enterprise Communications PCD-01 Communicate PCD Data message (also used for observations in response to a PCD-02 PCD Data Query) shall contain in MSH-21 Message Profile Identifier the value 1.3.6.1.4.1.19376.1.6.1.1.1
DEC-Vol2-010 For PCD-01 messages the ORC segment use is X, i.e. not sent.
DEC-Vol2-011 Pulse Oximetry Integration (POI) device observations shall include an observation of oxygen saturation of blood gas as a percentage numeric data type. Given the variety of available device designs (dedicated function SPO2 monitor or modular component to a physiologic monitor), the variety of sensor capabilities, and that the observation identification can include the sense point body part, i.e. left/right arm, etc. there is no single observation definition for all devices and all use cases. Any communicated SPO2 observation shall appear in RTMMS. A non-exclusive example is 150456^MDC_PULS_OXIM_SAT_O2^MDC.
PCD-Vol2-002 OBX-4 shall be expressed as the containment level of a particular item expressed in OBX-3
PCD-Vol2-003 Ordinal numbers in OBX-4 are not normative for a given parameter and may vary between implementations. They shall be unique within a given containment, however, numbers may change between messages.
PCD-Vol2-005 Periodic versus aperiodic data shall be distinguished by the IEEE 11073 Metric Access code in OBX-17
PCD-Vol2-006 The value in MSH-2 Encoding Characters shall be ^~\&
PCD-Vol2-007 MSH-3 Sending Application shall uniquely identify the software application implementing the actor sending the message
PCD-Vol2-008 If MSH-4 Sending Facility is populated, the first component shall identify the organizational entity responsible for sending the message, the second component if present shall identify the Universal ID of the organizational entity responsible for the sender of the message, the third component if present shall identify the Universal ID Type
PCD-Vol2-009 If MSH-5 Receiving Application is populated, the first component shall identify the organizational entity responsible for receiving the message, the second component if present shall identify the Universal ID of the organizational entity responsible for the sender of the message, the third component if present shall identify the Universal ID Type.
PCD-Vol2-010 If MSH-6 Receiving Facility is populated, the first component shall identify the organizational entity responsible for the receiving facility, the second component if present shall identify the Universal ID of the organizational entity responsible for the sender of the message, the third component if present shall identify the Universal ID Type
PCD-Vol2-011 In MSH-7 Date/Time of Message time zone shall be included in the field content
PCD-Vol2-012 In field MSH-11 Processing ID, the first component shall be present and be a value based upon HL7 Table 0103. The second component if present shall be a value based upon HL7 Table 0207.
PCD-Vol2-013 Field MSH-12 Version ID shall contain 2.6