Search Criteria : 37 assertions found for this search Review filtered assertions

Assertion

Applies to

Applied to
Not applied to

Coverage

Covered by
Not covered by
Id scheme
Assertion id
Status
Testable?
#Coverage
#Applies to
Comment
Predicate
Page
Tags
Last changed
Actions
LAB27LAB27-001validatedTestable 0 5 After the Analyzer working in query mode recognizes one or more specimens, the Analyzer sends a WOS Query Message (QBP^Q11^QBP_Q11) for each specimen to the Analyzer Manager.168Section 3.Q.53/7/16 2:23:29 PM by mpattillo
LAB27LAB27-005validatedTestable 0 1 An analyzer supporting the "Query All" Option SHALL be capable of sending a query for all pending work by using the query name WOS_ALL in field QPD-1. The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2. 171Section 3.Q.5.43/7/16 2:23:29 PM by mpattillo
LAB27LAB27-006validatedTestable 0 1 An analyzer supporting the "Query by Rack" Option SHALL be capable of sending a query identifying the position of the specimen in the rack (carrier). The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-4, QPD-5 and QPD-8. The first four are Mandatory, the latter is Required if Available. QPD-4 identifies the rack and QPD-5 is the position in the rack 171Section 3.Q.5.43/7/16 2:23:29 PM by mpattillo
LAB27LAB27-007validatedTestable 0 1 An analyzer supporting the "Query by Tray" Option SHALL be capable of sending a query identifying the position of the specimen in the tray. The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-6, QPD-7 and QPD-8. The first four are Mandatory, the latter is Required if Available. QPD-6 identifies the tray and QPD-7 is the position in the tray171Section 3.Q.5.43/7/16 2:23:29 PM by mpattillo
LAB27LAB27-008validatedTestable 0 1 An analyzer supporting the "Query by Isolate" Option SHALL be capable of sending a query identifying an isolate within the primary (parent) specimen. The only fields populated in the QPD segment SHALL be QPD-1 and QPD-2, QPD-3, and QPD-9. All four are Mandatory. QPD-3 identifies the isolate and QPD-9 identifies the primary specimen from which the microorganism was isolated. 171Section 3.Q.5.43/7/16 2:23:29 PM by mpattillo
LAB27LAB27-009validatedTestable 0 1 An analyzer supporting the "Bi-directional Communication" Option SHALL be capable of sending a query identifying the container. The only fields populated in the QPD segment SHALL be QPD-1, QPD-2, and QPD-3 (see Table X.5.1).171Section 3.Q.5.43/7/16 2:23:29 PM by mpattillo
LAB28LAB28-003validatedTestable 0 1 The Analyzer SHALL respond to the AWOS Broadcast message with an ORL^O24^ORL_O42 message compliant with table 3.R.5.2-2.180Section 3.R.5.13/7/16 2:23:29 PM by mpattillo
LAB28LAB28-004validatedTestable 0 1 If the OML message contains the Order Control Code NW, as part of the ORL^O34 response, the Analyzer SHALL transmit either Notification of request accepted or Unable to accept order/service for each AWOS contained in the OML message.185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB28LAB28-005validatedTestable 0 1 If the OML message contains the Order Control Code CA, as part of the ORL^O34 response, the Analyzer SHALL transmit either Cancel as requested or Unable to cancel for each AWOS contained in the OML message.185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB28LAB28-006validatedTestable 0 1 If the Analyzer Manager retransmits an AWOS or reuses an AWOS ID, the Analyzer SHALL reject the AWOS request by responding with Unable to accept order/service. 185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB28LAB28-008validatedTestable 0 5 A Negative Query Response for an unrecognized specimen is considered an unexpected situation, the Analyzer SHALL respond with an Application Acknowledgement: Reject in the ORL^O34 by setting MSA-1 to AR and reporting the appropriate specimen container values in the ERR segment.185Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB28LAB28-009validatedTestable 0 0 In response to a negative query response, the Analyzer shall send an ORL^OO34^ORL_O42 message containing only the MSH and MSA segments.191Section 3.R.5.33/7/16 2:23:29 PM by mpattillo
LAB29LAB29-001validatedTestable 0 1 The Analyzer SHALL notify the Analyzer Manager of the test results using the OUL^R22^OUL_R22 message.191Section 3.Y.5.311/18/16 5:24:28 PM by admin
LAB29LAB29-005validatedTestable 0 0 When more than one OBX segment is used to send related observations, different codes must be used in OBX-3 for each observation so that OBX-3 and OBX-4 are unique across OBXs.92Section W.2.5.23/7/16 2:23:29 PM by mpattillo
LAB29LAB29-006validatedTestable 0 0 The value of OBX-11 field across LAB-29 messages shall respect the Observation Result Status lifecycle for an Observation Run.135Figure W.3.6-13/7/16 2:23:29 PM by mpattillo
LAWLAW-001validatedTestable 0 1 Upon notification of AWOS completion by one of the Analyzers who transmits back the results, the Analyzer Manager shall cancel the other redundant AWOS awaiting execution on the other Analyzers.32Section X.2.13/7/16 9:59:00 AM by aberge
LAWLAW-002validatedTestable 0 0 In response to the AWOS broadcast, the Analyzer must notify the Analyzer Manager with its intent to accept or reject an AWOS by sending an ORL^O34^ORL_O42 message. 32Section X.2.1.13/7/16 9:59:37 AM by aberge
LAWLAW-004validatedTestable 0 0 If an Analyzer cannot cancel the AWOS, it should notify the Analyzer Manager that it is unable to cancel (using ORL^O34^ORL_O42 response message).34Section X.21.13/7/16 2:23:29 PM by mpattillo
LAWLAW-005validatedTestable 0 0 The AWOS ID is never entered manually on the Analyzer. It can only be obtained via a message from the Analyzer Manager. Therefore, when the AWOS is entered at the Analyzer, the AWOS shall have a value of NULL ("") for the null AWOS ID.36Section X.2.33/7/16 2:39:37 PM by aberge
LAWLAW-006validatedTestable 0 1 The Analyzer Manager must support receiving AWOS test results reported over a period of time in addition to receiving multiple test results for an AWOS at the same time.36Section X.2.33/7/16 10:13:57 AM by aberge